Reported Jul 6, 2016 · Terminated
Liquid stable 2 part Homocysteine Reagent, for in vitro quantitative determination of total homocysteine in human serum…
Medtest Holdings, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Medtest Holdings, Inc. · 5449 Research Dr · Canton, MI · 48188-2261
Is this the product you have?
- Liquid stable 2 part Homocysteine Reagent, for in vitro quantitative determination of total homocysteine in human serum and plasma - H7575-57 kit includes one Rl and one R2 bottle and one bottle of calibrator level 1 and one bottle of calibrator level 2.
- Lot codes and dates
- ModelH7575-57
- Lot416703
- Expiration date2016-01
- Amount recalled
- 13
Where it was sold
Why it was recalled
Deterioration in performance of the Liquid Stable (LS) 2-Part HOMOCYSTEINE REAGENT resulting in under recovery. The manufacturer has indicated that the HOMOCYSTEINE REAGENT lot listed above shows deterioration in assay performance manifesting itself by under recovery of Homocysteine in Controls and Patient Samples. The cause of this deterioration has been attributed to a loss of reductant activity in Reagent 1. Testing performed indicated that the affected lot is at the limit of acceptable clinical performance. This observation was consistent across the reportable range of the assay. For this reason, do not continue to use affected reagent lot. Test data generated to date will be acceptable where validity control specifiCations have been met.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.