Reported Jul 12, 2023 · Ongoing
OmniTom/OmniTom Elite- X-ray computed tomography applications for anatomy that can be imaged in the 40 cm aperture,…
NeuroLogica Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: NeuroLogica Corporation · 14 Electronics Ave · Danvers, MA · 01923-1011
Is this the product you have?
- OmniTom/OmniTom Elite- X-ray computed tomography applications for anatomy that can be imaged in the 40 cm aperture
- primarily head and neck Model NumberNL5000
- Lot codes and dates
- GTIN10815411020335 10815411020663
- Serial Numbers10 14 23 25 26 36 38 40 41 42 44 46 48 49 50 51 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 15R
- Amount recalled
- 48 units
Where it was sold
- Nationwide
Why it was recalled
1. Wheels loosening over time with use of the device could lead to the inability to effectively drive the device and could lead to delay in patient treatment. 2.Batteries used to power the NL5000 device contain contaminated boards, which may impact the ability to power on the device, this could lead to delay for patient treatment
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new NeuroLogica Corporation recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.