Reported Jul 23, 2014 · Terminated
Excelsior Medical Sterile Triple Lead Tubing Set and Sterile Heavy Duty Triple Lead Tubing Set. To be used with the…
Excelsior Medical Corp.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Excelsior Medical Corp · 1933 Heck Ave · Neptune, NJ · 07753
Is this the product you have?
- Excelsior Medical Sterile Triple Lead Tubing Set and Sterile Heavy Duty Triple Lead Tubing Set. To be used with the Pharm-Assist Dispensing Pump. Used for drug reconstitution and/or fluid transfer
- Lot codes and dates
Product Code PA03 - Lot Nos.: 0303081303050813 and 03091213 Product Code BX03Lot Nos.: 64031513640731136408131364090513
- Amount recalled
- 4250 tubing sets (PA03 - 78 cases; BX03 - 92 cases (each case contains 25 tubing sets))
Where it was sold
- Nationwide
Why it was recalled
There is a potential for a leak. A molding defect in the connector that joins the three leads to the main flow line may allow fluid to escape or air to enter the fluid path during transfer of fluids from one container to another.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Excelsior Medical Corp. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.