Reported Aug 1, 2012 · Ongoing

ABG II Modular Hip Stem Stryker Howmedica Osteonics Corp. Mahwah, NJ 07430 Stryker Ireland Carrigtwohill Industrial…

Stryker Howmedica Osteonics Corp.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Stryker Howmedica Osteonics Corp. · 325 Corporate Drive · Mahwah, NJ · 07430

Is this the product you have?

  • ABG II Modular Hip Stem Stryker Howmedica Osteonics Corp. Mahwah, NJ 07430 Stryker Ireland Carrigtwohill Industrial Estate, Carrigtwohill County Cork, Ireland. The ABG II Modular range of implants is constituted of 8 right stems and 8 left stems. The necks are made of GADS Vitallium, a proprietary Co Cr alloy that Stryker has developed. GADS stands for Gas Atomized Dispersion Strengthened. IT is a Co Cr alloy that is a corrosion resistant and demonstrates qualities of improved fatigue strength (with respect to regular Co Cr alloy). Neck Range - The ABG II modular neck range is composed of 10 reversible neck implants. There are different options in terms of length, version and neck angle. The AG II Modular
  • necks come in two different lengths28 mm (short) and 36 mm (long). The necks are offered in two version angles ; 0 and 7. The 0 and 7 necks are available in 125, 130 and 135 neck angles
Lot codes and dates
  • 510 K092406 all lots (4845-4XXX and NLS/NLV)
Amount recalled
9003 units

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

  • Nationwide

Why it was recalled

Stryker has updated the instruction for use (IFU) (product correction)for the ABGII Modular and Rejuvenate Modular Hip Systems. This is based on a reported rate of less than one percent for revisions potentially associated with fretting and or corrosion at or about the modular neck junction. On June 29, 2012, Stryker has decided to remove the affected product from the field since there may be an

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Stryker Howmedica Osteonics Corp. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:24 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.