Reported Jul 29, 2015 · Terminated

0702-001-350Q, Diverter Valve Assembly 0702-001-000, Neptune 2 Rover Ultra (120V) 0702-001-000R, Repair Neptune 2 Rover…

Stryker Instruments Div. of Stryker Corporation

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Stryker Instruments Div. of Stryker Corporation · 4100 E Milham Ave · Portage, MI · 49002-9704

Is this the product you have?

  • 0702-001-350Q, Diverter Valve Assembly 0702-001-000, Neptune 2 Rover Ultra (120V) 0702-001-000R, Repair Neptune 2 Rover 0702-002-000, Neptune 2 Rover Ultra (230V) 0702-002-000ES, Neptune 2 Rover (230V) Spanish 0702-002-000FI, Neptune 2 Ultra Rover (230V) Finnish 0702-002-000IT, Neptune 2 Rover (230V) Italian 0702-002-000R, Repair Neptune 2
  • Rover Ultra (230V) Product UsageThe Neptune 2 Waste Management System is intended to be used in the operating room, pathology, surgical centers, and doctor s offices to collect and dispose of surgical fluid waste as well as collect smoke generated from electro cautery or laser devices
  • Contraindicationconnection of the Neptune 2 Waste Management System directly to chest tubes or other closed wound drains. The Stryker Neptune 2 Rover is a mobile unit used to suction, collect fluid waste, and evacuate surgical smoke from a surgical site during operation. Fluid waste is removed from the surgical site through suction tubing connected to inlet ports of disposable manifolds installed in the rover. The Diverter Valve Assembly consists of two diverter valves with three hoses, one inlet and two outlets. The sole purpose of the Diverter Valve Assembly is to direct/divert water from the fresh water inlet to either the large canister or small canister via the pre-fill tank
Lot codes and dates
  • Serial Numbers ranging from 0803118093 through 1505400223
  • Distribution or Repair dates
    • November 1
    • 2011 to March 31
    • 2015
Amount recalled
9,245

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

Why it was recalled

Stryker Instruments is initiating a recall of the Neptune 2 Waste Management System due to complaints related to installation failures of the Neptune 2 diverter valve assembly during docking.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Stryker Instruments Div. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Sep 12, 2016 · Data as of Sep 19, 5:24 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.