Reported Jul 12, 2023 · Ongoing

BD Synapsys Informatics Solution, Catalog Number 444150

Becton Dickinson & Co.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Becton Dickinson & Co. · 7 Loveton Cir · Sparks, MD · 21152-9212

Is this the product you have?

  • BD Synapsys Informatics Solution, Catalog Number 444150
Lot codes and dates
  • UDI-DI00382904441500
Amount recalled
102 systems

Where it was sold

Why it was recalled

There is a risk of mis-association of patient demographic data for users who choose to report results manually when their Service Control Unit (SCU) is operating near capacity and are using save actions (Skip, Final Discard, Mark as Read, Save) to process patient results rapidly.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Becton Dickinson & Co. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:38 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.