Reported Jul 12, 2023 · Ongoing
BD Synapsys Informatics Solution, Catalog Number 444150
Becton Dickinson & Co.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Becton Dickinson & Co. · 7 Loveton Cir · Sparks, MD · 21152-9212
Is this the product you have?
- BD Synapsys Informatics Solution, Catalog Number 444150
- Lot codes and dates
- UDI-DI00382904441500
- Amount recalled
- 102 systems
Where it was sold
Why it was recalled
There is a risk of mis-association of patient demographic data for users who choose to report results manually when their Service Control Unit (SCU) is operating near capacity and are using save actions (Skip, Final Discard, Mark as Read, Save) to process patient results rapidly.