Reported Aug 14, 2019 · Terminated
YXLON Cheetah cabinet X-Ray System Product Usage: The products are generally used for non-destructive testing and…
Comet Technologies USA Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Comet Technologies Usa Inc · 100 Trap Falls Rd · Shelton, CT · 06484-4604
Is this the product you have?
- YXLON Cheetah
- cabinet X-Ray System Product UsageThe products are generally used for non-destructive testing and quality control
- Lot codes and dates
- X-Ray Systems
- Amount recalled
- 77
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
- Nationwide
Why it was recalled
Failure to meet the requirement of the performance standard for cabinet x-ray systems (21 CFR 1020.40) for a door interlock that causes physical disconnection of the power supply using only the movement of the door.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Comet Technologies USA Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.