Reported Aug 8, 2012 · Terminated

Maquet Getinge Group Operating Table System Magnus Carbon-Fibre Table Top Version 1180.16 The carbon-fibre MAGNUS table…

Maquet Medical Systems, USA

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Maquet Medical Systems, Usa · 45 Barbour Pond Dr · Wayne, NJ · 07470-2094

Is this the product you have?

  • Maquet Getinge Group Operating Table System Magnus Carbon-Fibre Table Top Version 1180.16 The carbon-fibre MAGNUS table top (1180-16XX) serves to support and position patients for surgical procedures (e.g. cardiovascular surgery) and interventional examination immediately before, during and after the surgical phase. The table top is radiotranslucent and enables intraoperative use of X-ray equipment.
Lot codes and dates
    • Model # 1180.16F2 Serial #s 1
    • 3
    • 7
    • & 8
    • Model # 1180.16F3 Serial # 1
    • Model #1180.16F5 Serial #4
    • Model # T285.7000 Serial #s 1
    • 2
    • 3
    • 4
    • & 7
Amount recalled
164 Units located in 27 Countries

Where it was sold

  • Nationwide

Why it was recalled

As a result of a small number of customer complaints, MAQUET conducted an investigation and identified a potential problem with specific MAGNUS Carbon-Fibre Table Tops manufactured between February 23, 2010 and February 28, 2012 and distributed between November 11, 2010 and April 6, 2012. There is a potential issue with the motor brakes for the longitudinal shift drive, and the tilt drive.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Maquet Medical Systems, USA recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Sep 4, 2013 · Data as of Sep 19, 5:24 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.