Reported Jun 7, 2017 · Terminated
Angio WorkStation (XIDF-AWS801) used in conjunction with your Infinix System (INFX-8000V;INFX-8000C;INFX-8000F
Toshiba American Medical Systems Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Toshiba American Medical Systems Inc · PO Box 2068 · Tustin, CA · 92780-7047
Is this the product you have?
- Angio WorkStation (XIDF-AWS801) used in conjunction with your Infinix System (INFX-8000V;INFX-8000C;INFX-8000F
- Lot codes and dates
- S/N8181622393
- Amount recalled
- US - 70 systems
Where it was sold
Why it was recalled
It was found that during a procedure the Peak Skin Dose (PSD) value displayed by the Dose Tracking System (DTS) was larger than the DTS expected value. This issue occurs when the power for the Angio Workstation is turned on after the power for the Infinix system is turned on.