Reported Jun 13, 2018 · Terminated
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 6), infusion pump, 35700BAX and 35700ABB Product…
Baxter Healthcare Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Baxter Healthcare Corporation · 711 Park Ave · Medina, NY · 14103-1036
Is this the product you have?
- Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 6), infusion pump, 35700BAX
- and 35700ABB Product UsageThe Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used for the controlled administration of fluids. These fluids may include pharmaceutical drugs, blood, blood products and mixture of required patient therapy. The intended routes of administration consist of the
- following clinically accepted routesintravenous, arterial, subcutaneous, intrathecal, epidural or irrigation of fluid space. Spectrum Infusion Pump is intended to be used in conjunction with legally marketed intravenous administration sets and medications provided by the user. The Spectrum and Spectrum with Master Drug Library is suitable for many user facility applications such as but not limited to hospitals, outpatient care areas, homecare and ambulatory care services. Spectrum and Spectrum with Master Drug Library is intended to reduce operator interaction through automated programming thereby helping to reduce errors associated with complex device programming. Parameter programming requires trained healthcare professional confirmation of limits and drug therapy to the physicians directive
- Lot codes and dates
- All serial numbers manufactured from 11/1/2005 to 4/22/2013, with release dates 1/1/2005 and forward.
- Amount recalled
- 333961
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
- Nationwide
Why it was recalled
The firm will be updating the Instructions for Use for the product. Current IFUs instruct users to lower the primary bag 20 inches below the secondary bag using a hanger when performing a secondary infusion. However, the firm supplies 12.5-inch hangers with the Baxter disposable secondary sets.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Baxter Healthcare Corporation recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.