Reported Jul 23, 2025 · Ongoing
SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve, REF: SCXT3-5-2000; SwabCap, 200-ct. Box, REF: SCXT3-2000
ICU Medical, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: ICU Medical, Inc. · 951 Calle Amanecer · San Clemente, CA · 92673-6212
Is this the product you have?
- SwabCap Disinfecting
- Cap for Needlefree ConnectorsSwabSleeve
- REF
SCXT3-5-2000SwabCap200-ct. Box
- REFSCXT3-2000
- Lot codes and dates
2 of 2 products · 31 lot codes
Recalled lot codes, grouped by product number REF UDI-DI Lot (Expiration) SCXT3-5-200010887709097107142587391-Dec-26142587431-Feb-26142587441-Dec-26142642001-Dec-26142722411-Dec-26142722481-Dec-26142874561-Dec-26142874591-Dec-26142874801-Dec-26142874821-Dec-26142874851-Dec-26142948321-Dec-26142722541-Dec-26143103131-Dec-26143104251-Dec-26143233541-Dec-26142874751-Dec-26142948401-Dec-26143216451-Jan-27143216521-Jan-27143311691-Jan-27143327731-Jan-27143417321-Oct-26142948251-Jan-27143032091-Jan-27143103081-Jan-27143554041-Jan-27143554091-Jan-27
SCXT3-200020887709079162142582141-Jan-27142582061-Jan-27142582081-Feb-27
- Amount recalled
- 1,445,600
Where it was sold
Why it was recalled
Due to a manufacturing issue, disinfecting cap for needle-free connectors may have an incomplete seal between the foil lid and plastic container, which may result in isopropyl alcohol evaporation from the sponge, which may result in inadequate disinfection.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new ICU Medical, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.