Reported Aug 8, 2012 · Terminated
Equinoxe Reverse Compression Screw Intended for Primary and Reverse Total Shoulder Arthroplasty.
Exactech, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Exactech, Inc. · 2320 NW 66th Ct · Gainesville, FL · 32653-1630
Is this the product you have?
- Equinoxe Reverse Compression Screw Intended for Primary and Reverse Total Shoulder Arthroplasty.
- Lot codes and dates
- Sizes4.5x18mm Catalog # 320-20-18 Serial Range 1610445-2381151 4.5x22mm Catalog # 320-20-22 Serial Range 1610391-2381151 4.5x26mm Catalog # 320-20-26 Serial Range 1611237-2381151 4.5x30mm Catalog # 320-20-30 Serial Range 1617556-2381151 4.5x34mm Catalog # 320-20-34 Serial Range 1611290-2381151 4.5x38mm Catalog # 320-20-38 Serial Range 1610260-2381151 4.5x42mm Catalog # 320-20-42 Serial Range 1630006-2381151 4.5x46mm Catalog # 320-20-46 Serial Range 1644317-2381151
- Amount recalled
- 54,877 units for all products in Recall Event
This recall covers 6 products
The agency filed one notice per product. You are reading one of them; the other 5 are below.
Where it was sold
Why it was recalled
Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Exactech, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.