Reported Jul 23, 2025 · Ongoing
VITROS Chemistry Products CHE Slides; Catalog Numbers: (1) 191 4605 (300 slides), (2) 800 4707 (90 slides);
Ortho-Clinical Diagnostics, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Ortho-Clinical Diagnostics, Inc. · 100 Indigo Creek Dr · Rochester, NY · 14626-5101
Is this the product you have?
- VITROS Chemistry Products CHE Slides
- Catalog Numbers
(1) 191 4605 (300 slides)(2) 800 4707 (90 slides)
- Lot codes and dates
- Catalog Numbers
(1) 191 4605 (300 slides)(2) 800 4707 (90 slides)
- UDI-DI
(1) 10758750000241(2) 10758750004225
- Lot NumbersAll lots within expiry
- Catalog Numbers
- Amount recalled
- 27, 330 US/ 19,679,730 OUS
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Where it was sold
Why it was recalled
Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Ortho-Clinical Diagnostics, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.