Reported May 31, 2017 · Terminated
Merge Cardio software. The firm name on the label is Merge Healthcare, Hartland, WI. Intended to be used to acquire,…
Merge Healthcare, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Merge Healthcare, Inc. · 900 Walnut Ridge Dr · Hartland, WI · 53029-8347
Is this the product you have?
- Merge Cardio software. The firm name on the label is Merge Healthcare, Hartland, WI. Intended to be used to acquire, store, print, transfer, and archive clinical information.
- Lot codes and dates
Versions 10.010.0.110.110.1.110.1.1.1and 10.1.2
- Amount recalled
- 30 sites have the potentially affected software versions
Where it was sold
Why it was recalled
Fetal patient report was automatically pulling prior measurement data for a prior fetus since the fetal study is based on the mother's MRN, resulting in the wrong fetal measurements getting referenced.
This product could cause temporary or reversible health problems — stop using it.