Reported May 31, 2017 · Terminated

Microalbumin Reagent, Catalog No. 475100, Product Labeling A18520AL MA reagent, when used in conjunction with UniCel…

Beckman Coulter Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Beckman Coulter Inc. · 250 S Kraemer Blvd · Brea, CA · 92821-6232

Is this the product you have?

  • Microalbumin Reagent, Catalog No. 475100, Product Labeling A18520AL MA reagent, when used in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON¿ Systems MA Calibrator, is intended for quantitative determination of Albumin concentration in human urine
Lot codes and dates
  • M609458
Amount recalled
3,158 units total (1,957 units in US)

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

  • Nationwide

Why it was recalled

Beckman Coulter has confirmed that MA Reagent Lot M609458 may recover outside the lower limit of the published insert range for Beckman Coulter Urine Protein Control Level 1.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Beckman Coulter Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Dec 30, 2017 · Data as of Sep 19, 6:00 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.