Reported Aug 8, 2012 · Terminated
Flush Kit w 3 Way Off Stopcock Product Usage: Flush Kit
ICU Medical, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: ICU Medical, Inc. · 951 Calle Amanecer · San Clemente, CA · 92673-6212
Is this the product you have?
- Flush
- Kit w 3 Way Off Stopcock Product UsageFlush Kit
- Lot codes and dates
- Part Number AG8137 510(k) K052865 Lot number 2466161
- Amount recalled
- N/A
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Where it was sold
- Nationwide
Why it was recalled
The firm initiated the recall due to a molding defect in a high pressure stopcock where a small percentage of the side ports could become totally occluded.