Reported Jun 3, 2020 · Ongoing

da Vinci SP Camera; PN 430060-33; Model Number: 430060-33. For endoscopic surgery.

Intuitive Surgical, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Intuitive Surgical, Inc. · 1266 Kifer Rd Bldg 101 · Sunnyvale, CA · 94086-5304

Is this the product you have?

  • da Vinci SP Camera; PN 430060-33
  • Model Number430060-33. For endoscopic surgery
Lot codes and dates
  • Model Number430060-33
  • UDI Number00886874114384; All serial numbers
Amount recalled
305 units

Where it was sold

  • Nationwide

Why it was recalled

Some cameras were found to have residual soil adhered to the tip of the camera after reprocessing which may affect image quality resulting in needing to switch to a second camera or converting to another surgical modality. If residual soil makes contact with the patient, exposure could result in sensitization making future component blood therapy or tissue typing more challenging; additionally if a small piece of inert epoxy were to be retained, scar tissue could form.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Intuitive Surgical, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:26 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.