Reported Aug 8, 2012 · Terminated
I-STAT EG7+ Cartridge Abbott Point of Care, Princeton, NJ. The i-Stat EG7 cartridge is intended to be used for…
Abbott Point Of Care Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Abbott Point Of Care Inc. · 400 College Rd E · Princeton, NJ · 08540-6607
Is this the product you have?
- I-STAT EG7+ Cartridge Abbott Point of Care, Princeton, NJ. The i-Stat EG7 cartridge is intended to be used for quantitatively measure sodium, potassium, ionized calcium, oxygen, carbon dioxide, partial pressure in arterial, venous, or capillary whole blood.
- Lot codes and dates
- List number 03P76-25 Lot number S10246A
- Amount recalled
- 52,750 cartridges
Where it was sold
- Nationwide
Why it was recalled
i-Stat EG7 cartridges may generate falsely depressed PCO2 results for samples above 26 mm hg when stored at room temperature for four weeks or longer.
This product could cause temporary or reversible health problems — stop using it.