Reported Jul 30, 2025 · Ongoing
SIGNA PET/MR
GE Medical Systems, LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: GE Medical Systems, LLC · 3200 N Grandview Blvd · Waukesha, WI · 53188-1693
Is this the product you have?
- SIGNA PET/MR
- Lot codes and dates
- GTIN00840682105378
- Amount recalled
- N/A
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Where it was sold
- Nationwide
Why it was recalled
GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.
This product could cause temporary or reversible health problems — stop using it.