Reported Jun 3, 2020 · Terminated

ABL90 FLEX PLUS, model no. 393-02; and Spare Part Hemolyzer Units (902-980): R0526 - R0941 - Product Usage: are…

Radiometer Medical ApS

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Radiometer Medical ApS · Akandevej 21 · Bronshoj, N/A · N/A

Is this the product you have?

    • ABL90 FLEX PLUS
    • model no. 393-02
    • and
  • Spare Part Hemolyzer Units (902-980)R0526 -
  • R0941 - Product Usageare intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting
Lot codes and dates
  • ABL90 FLEX PLUS Model no. 393-092: R0014 - R0195; and
  • Spare Part Hemolyzer Units (902-980)
    • R0526 - R0941
    • Model Number 393-092
Amount recalled
544

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

  • Nationwide

Why it was recalled

The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted through the sample. This will lead to a perturbed and non-linear absorbance spectrum which results in measurement errors.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Radiometer Medical ApS recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Nov 8, 2024 · Data as of Sep 19, 6:14 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.