Reported Jul 29, 2015 · Terminated
ATTUNE CONV RP PS ARTICULATION SURFACE SZ1-9 INTENDED USE: The Tibial Articulation Surface Trial snaps together with…
DePuy Orthopaedics, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: DePuy Orthopaedics, Inc. · 700 Orthopaedic Dr · Warsaw, IN · 46582-3994
Is this the product you have?
- ATTUNE CONV RP PS
- ARTICULATION SURFACE SZ1-9 INTENDED USEThe Tibial Articulation Surface Trial snaps together with the Shim component to function as the Insert Trial during a total knee replacement. A Balseal is attached to each of the two post features on the Articulation Surface to provide a connection force between the Tibial Articulation Surface and Shim to ensure secure engagement between the components
- Lot codes and dates
- Product Codes
254500991254500992254500993254500994254500995254500996254500997254500998and 254500999
- GTIN
1060329513568510603295135692106032951357081060329513571510603295135722106032951357391060329513574610603295135753and 10603295135760
- Product Codes
- Amount recalled
- 11360 units
This recall covers 10 products
The agency filed one notice per product. You are reading one of them; the other 9 are below.
Where it was sold
Why it was recalled
Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal is separated from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new DePuy Orthopaedics, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.