Reported Jul 26, 2023 · Ongoing
Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-31 0998-00-0800-32 0998-00-0800-33…
Datascope Corp.
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Datascope Corp. · 1300 Macarthur Blvd · Mahwah, NJ · 07430-2052
Is this the product you have?
- Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model numbers0998-UC-0800-31* 0998-UC-0800-33* 0998-UC-0800-52* 0998-UC-0800-53* 0998-00-0800-55* *Not US distribution
0998-00-0800-310998-00-0800-320998-00-0800-330998-00-0800-340998-00-0800-350998-00-0800-450998-00-0800-520998-00-0800-530998-00-0800-550998-00-0800-65
- Lot codes and dates
- All Unit Serial Numbers Model UDI
0998-00-0800-31106075671090530998-00-0800-32106075671111170998-00-0800-33106075671090080998-00-0800-34106075671119400998-00-0800-35106075671091070998-00-0800-45106075671084210998-00-0800-52106075671084380998-00-0800-53106075671083910998-00-0800-55106075671084140998-00-0800-6510607567113432
- All Unit Serial Numbers Model UDI
- Amount recalled
- 9174 (4586 US; 4588 OUS)
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Where it was sold
- Nationwide
Why it was recalled
Stop using this product now
IABP may lose the ability to charge batteries in one or both bay slots. Therapy may be interrupted if batteries fail to charge and the device is disconnected from AC power. Low battery alarms may alert the User to the issue prior to interruption of therapy.
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Datascope Corp. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.