Reported Aug 13, 2025 · Ongoing

BD Alaris Pump Infusion Sets REF: 10013186, BD Alaris Pump Infusion Set, Back Check Valve, 3 SmartSite Y-sites LVP 60D…

CareFusion 303, Inc.

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: CareFusion 303, Inc. · 10020 Pacific Mesa Blvd · San Diego, CA · 92121-4386

Is this the product you have?

16 of 16 products

Products covered by this recall
Model / numberProduct as published
10013186BD Alaris Pump Infusion Sets , BD Alaris Pump Infusion Set, Back Check Valve, 3 SmartSite Y-sites LVP 60D DEHP FREE 3SS CV
2452-0007BD Alaris Pump Infusion Set, 2 Back Check Valves, 3 SmartSite Y-sitesAS LVP 10D 3SS 2CV
2204-0007BD Alaris Pump Infusion SetAS LVP 20D
24001-0007BD Alaris Pump Infusion Set, Back Check Valve, SmartSite Y-siteAS LVP 20D SS CV
10015012BD Alaris Pump Infusion Burette Set, 0.2 Micron Filter, Smallbore Tubing, SmartSite Port (Burette), SmartSite Y-siteAS LVP BUR 20D PE-Lined SMBORE SS 0.2M
2432-0007BD Alaris Pump Infusion Set, 0.2 Micron Filter, Back Check Valve, 3 SmartSite Y-sitesAS LVP 20D 3SS 0.2M CV
2434-0007BD Alaris Pump Infusion Set, 0.2 Micron Filter, Back Check Valve, SmartSite Y-siteAS LVP 20D SS 0.2M CV
2202-0007BD Alaris Pump Infusion Set, 1.2 Micron FilterAS LVP 20D 1.2M
11426965BD Alaris Pump Infusion Set, Back Check Valve, 5 SmartSite Y-sitesAS LVP 20D CV
10561554BD Alaris Pump Infusion Set, 15 Micron Filter, Back Check Valve, 3 SmartSite Y-sitesAS LVP 20D DEHP FREE 3SS 15M CV
10863358BD Alaris Pump Infusion Set, 15 Micron FilterAS LVP 20D DEHP FREE 15M
11171447BD Alaris Pump Infusion Set, 2 Back Check Valves, 3 SmartSite Y-sitesAS LVP 20D 3SS 2CV
11532269BD Alaris Pump Infusion Set, 0.2 Micron Filter, Back Check Valve, PE Lined Tubing, 2 SmartSite Y-sitesAS LVP 20D PE-LINED 2SS 0.2M CV
10010454BD Alaris Pump Infusion Set, 0.2 Micron Filter, PE Lined Tubing, SmartSite Y-siteAS LVP 20D PE-LINED TUBING SS 0.2M
2465-0007BD Alaris Pump Infusion Set, SmartSite Bag Access Non-Vented, 0.2 Micron Filter, Low Sorbing Tubing (PE Lined), SmartSite Y-siteAS LVP SS BAG 20D PE-Lined SS 0.2M
24301-0007TREF: 10010453, ASBD Alaris Pump Infusion Set, 1.2 Micron Filter, PE Lined Tubing, SmartSite Y-site LVP 20D PE-LINED TUBING SS 1.2M REF: 11419365, BD Alaris" Pump Infusion Set, Back Check Valve, Manifold, 3-Way Stopcock, 5 SmartSite" Y-sites REF: 10813621, BD Alaris" Pump Infusion Set, Back Check Valve, 3 Ganged 4-Way Stopcocks, 4 SmartSite" Y-sites REF: 2423-0007, BD Alaris" Pump Infusion Set, Back Check Valve, 2 Ganged 4-Way Stopcocks, 4 SmartSite" Y-sites REF: 10015896, BD Alaris" Pump Infusion Set, Back Check Valve, 2 Ganged 4-Way Stopcocks, 3 SmartSite" Y-sites REF: 24010-0007T, BD Alaris" Pump Infusion Set, Bonded Texium", Closed Male Luer with Priming Cap, Back Check Valve, 3 SmartSite" Y-sites REF: 10010483, BD Alaris" Pump Infusion Set, Vented Syringe Adapter, Smallbore Tubing REF: 2403-0007, BD Alaris" Pump Infusion Set, Half Set, SmartSite" Y-site , BD Alaris" Pump Infusion Set, SmartSite" Bag Access Non-Vented, 0.2 Micron Filter, PE Lined Tubing, Bonded Texium" Closed Male Luer with Priming Cap, SmartSite" Y-site
Lot codes and dates

16 of 16 products

Products covered by this recall
Model / numberBarcodesProduct as published
10013186/07613203021173UDI: 07613203021173
2452-0007/10885403219870UDI: 10885403219870
2204-0007/10885403199363UDI: 10885403199363
24001-0007/10885403238666UDI: 10885403238666
10015012/10885403233951UDI: 10885403233951
2432-0007/10885403232329UDI: 10885403232329
2434-0007/07613203019682UDI: 07613203019682
2202-0007/10885403274039UDI: 10885403274039
11426965/10885403232558UDI: 10885403232558
10561554/10885403232565UDI: 10885403232565
10863358/10885403232466UDI: 10885403232466
11171447/07613203021234UDI: 07613203021234
11532269/10885403232343UDI: 10885403232343
10010454/07613203015806UDI: 07613203015806
2465-0007/10885403221941UDI: 10885403221941
24301-0007T/UDI076132030211351088540323473610885403234514076132030209851088540323406410885403271021108854032326021088540323247310885403223198REF: 10010453/ UDI: 07613203021135 REF: 11419365/ UDI: 10885403234736 REF: 10813621/UDI: 10885403234514 REF: 2423-0007/UDI: 07613203020985 REF: 10015896/UDI: 10885403234064 REF: 24010-0007T/UDI: 10885403271021 REF: 10010483/UDI: 10885403232602 REF: 2403-0007/UDI: 10885403232473 : 10885403223198
Amount recalled
33,162,680 units total

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

Why it was recalled

Stop using this product now

Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new CareFusion 303, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:54 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.