Reported Aug 13, 2025 · Ongoing
BD Alaris Pump Infusion Sets REF: 10013186, BD Alaris Pump Infusion Set, Back Check Valve, 3 SmartSite Y-sites LVP 60D…
CareFusion 303, Inc.
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: CareFusion 303, Inc. · 10020 Pacific Mesa Blvd · San Diego, CA · 92121-4386
Is this the product you have?
16 of 16 products
| Model / number | Product as published |
|---|---|
10013186 | BD Alaris Pump Infusion Sets , BD Alaris Pump Infusion Set, Back Check Valve, 3 SmartSite Y-sites LVP 60D DEHP FREE 3SS CV |
2452-0007 | BD Alaris Pump Infusion Set, 2 Back Check Valves, 3 SmartSite Y-sitesAS LVP 10D 3SS 2CV |
2204-0007 | BD Alaris Pump Infusion SetAS LVP 20D |
24001-0007 | BD Alaris Pump Infusion Set, Back Check Valve, SmartSite Y-siteAS LVP 20D SS CV |
10015012 | BD Alaris Pump Infusion Burette Set, 0.2 Micron Filter, Smallbore Tubing, SmartSite Port (Burette), SmartSite Y-siteAS LVP BUR 20D PE-Lined SMBORE SS 0.2M |
2432-0007 | BD Alaris Pump Infusion Set, 0.2 Micron Filter, Back Check Valve, 3 SmartSite Y-sitesAS LVP 20D 3SS 0.2M CV |
2434-0007 | BD Alaris Pump Infusion Set, 0.2 Micron Filter, Back Check Valve, SmartSite Y-siteAS LVP 20D SS 0.2M CV |
2202-0007 | BD Alaris Pump Infusion Set, 1.2 Micron FilterAS LVP 20D 1.2M |
11426965 | BD Alaris Pump Infusion Set, Back Check Valve, 5 SmartSite Y-sitesAS LVP 20D CV |
10561554 | BD Alaris Pump Infusion Set, 15 Micron Filter, Back Check Valve, 3 SmartSite Y-sitesAS LVP 20D DEHP FREE 3SS 15M CV |
10863358 | BD Alaris Pump Infusion Set, 15 Micron FilterAS LVP 20D DEHP FREE 15M |
11171447 | BD Alaris Pump Infusion Set, 2 Back Check Valves, 3 SmartSite Y-sitesAS LVP 20D 3SS 2CV |
11532269 | BD Alaris Pump Infusion Set, 0.2 Micron Filter, Back Check Valve, PE Lined Tubing, 2 SmartSite Y-sitesAS LVP 20D PE-LINED 2SS 0.2M CV |
10010454 | BD Alaris Pump Infusion Set, 0.2 Micron Filter, PE Lined Tubing, SmartSite Y-siteAS LVP 20D PE-LINED TUBING SS 0.2M |
2465-0007 | BD Alaris Pump Infusion Set, SmartSite Bag Access Non-Vented, 0.2 Micron Filter, Low Sorbing Tubing (PE Lined), SmartSite Y-siteAS LVP SS BAG 20D PE-Lined SS 0.2M |
24301-0007T | REF: 10010453, ASBD Alaris Pump Infusion Set, 1.2 Micron Filter, PE Lined Tubing, SmartSite Y-site LVP 20D PE-LINED TUBING SS 1.2M REF: 11419365, BD Alaris" Pump Infusion Set, Back Check Valve, Manifold, 3-Way Stopcock, 5 SmartSite" Y-sites REF: 10813621, BD Alaris" Pump Infusion Set, Back Check Valve, 3 Ganged 4-Way Stopcocks, 4 SmartSite" Y-sites REF: 2423-0007, BD Alaris" Pump Infusion Set, Back Check Valve, 2 Ganged 4-Way Stopcocks, 4 SmartSite" Y-sites REF: 10015896, BD Alaris" Pump Infusion Set, Back Check Valve, 2 Ganged 4-Way Stopcocks, 3 SmartSite" Y-sites REF: 24010-0007T, BD Alaris" Pump Infusion Set, Bonded Texium", Closed Male Luer with Priming Cap, Back Check Valve, 3 SmartSite" Y-sites REF: 10010483, BD Alaris" Pump Infusion Set, Vented Syringe Adapter, Smallbore Tubing REF: 2403-0007, BD Alaris" Pump Infusion Set, Half Set, SmartSite" Y-site , BD Alaris" Pump Infusion Set, SmartSite" Bag Access Non-Vented, 0.2 Micron Filter, PE Lined Tubing, Bonded Texium" Closed Male Luer with Priming Cap, SmartSite" Y-site |
- Lot codes and dates
16 of 16 products
Products covered by this recall Model / number Barcodes Product as published 10013186/07613203021173UDI: 07613203021173 2452-0007/10885403219870UDI: 10885403219870 2204-0007/10885403199363UDI: 10885403199363 24001-0007/10885403238666UDI: 10885403238666 10015012/10885403233951UDI: 10885403233951 2432-0007/10885403232329UDI: 10885403232329 2434-0007/07613203019682UDI: 07613203019682 2202-0007/10885403274039UDI: 10885403274039 11426965/10885403232558UDI: 10885403232558 10561554/10885403232565UDI: 10885403232565 10863358/10885403232466UDI: 10885403232466 11171447/07613203021234UDI: 07613203021234 11532269/10885403232343UDI: 10885403232343 10010454/07613203015806UDI: 07613203015806 2465-0007/10885403221941UDI: 10885403221941 24301-0007T/UDI076132030211351088540323473610885403234514076132030209851088540323406410885403271021108854032326021088540323247310885403223198REF: 10010453/ UDI: 07613203021135 REF: 11419365/ UDI: 10885403234736 REF: 10813621/UDI: 10885403234514 REF: 2423-0007/UDI: 07613203020985 REF: 10015896/UDI: 10885403234064 REF: 24010-0007T/UDI: 10885403271021 REF: 10010483/UDI: 10885403232602 REF: 2403-0007/UDI: 10885403232473 : 10885403223198 - Amount recalled
- 33,162,680 units total
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
Why it was recalled
Stop using this product now
Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new CareFusion 303, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.