Reported Aug 20, 2014 · Terminated

VITROS¿ Immunodiagnostic Products CA 19-9 Calibrators and Reagent Packs, Lot 1320 VITROS¿ CA 19-9 Reagent Packs are…

Ortho-Clinical Diagnostics

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: ORTHO-CLINICAL DIAGNOSTICS · FELINDRE MEADOWS · Bridgend, N/A · N/A

Is this the product you have?

  • VITROS¿ Immunodiagnostic Products CA 19-9 Calibrators and Reagent Packs, Lot 1320 VITROS¿ CA 19-9 Reagent Packs are used on VITROS¿ ECi/ECiQ Immunodiagnostic Systems, VITROS¿ 3600 Immunodiagnostic Systems and VITROS¿ 5600 Integrated Systems for the quantitative measurement of 1116-NS-19-9 defined antigen in human serum and plasma.
Lot codes and dates
  • Lot 1320
  • Product code6800035 and 6800040
Amount recalled
475

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

  • Nationwide

Why it was recalled

Customers may receive positively biased results when using VITROS¿ Immunodiagnostic Products CA 19-9 Total Reagent Packs, Lot 1320 and VITROS¿ CA 19-9 Calibrators, Lot 1320 on the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System. There is an unexpected observed change in CA 19-9 Calibrators, Lot 1320 calibrator values.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Ortho-Clinical Diagnostics recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Mar 6, 2017 · Data as of Sep 19, 5:19 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.