Reported Aug 11, 2021 · Ongoing
REF: 08P1820, Abbott Alinity c Activated Alanine Aminotransferase reagent Kit, Alinity c Activated Alanine…
Abbott Laboratories
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Abbott Laboratories · 1915 Hurd Dr · Irving, TX · 75038
Is this the product you have?
- REF08P1820, Abbott Alinity c Activated Alanine Aminotransferase reagent Kit, Alinity c Activated Alanine Aminotransferase Reagent Kit,, IVD, CE
- UDI(01) 00380740132569
- Lot codes and dates
- Lot Numbers / Expiration Date 83459UN20 /09JUL2021, 03168UN20 /25SEP2021, 09787UN20 /24OCT2021, 37977UN20 /01JAN2022, 63000UN20 /01JAN2022, 77745UN20 /08JAN2022, 77743UN21 /02MAR2022, 09237UN21 /09APR2022
- Amount recalled
- 4,727 units
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Where it was sold
Why it was recalled
There is a potential for 15% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with samples containing bilirubin concentrations greater that 31 mg/dL. This could lead to incorrect patient results.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Abbott Laboratories recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.