Reported Jul 3, 2024 · Ongoing
Infinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Rev 20, 21, and 23
Draeger Medical Systems, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Draeger Medical Systems, Inc. · 6 Tech Dr · Andover, MA · 01810-2434
Is this the product you have?
- Infinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Rev 20, 21, and 23
- Lot codes and dates
- Model NumberMS20407
- UDI-DI
04049098054447. Rev 2021and 23
- Amount recalled
- 34,102 units
Where it was sold
- Nationwide
- GA
Why it was recalled
Device is not in full compliance to Type CF requirements of IEC 60601-1 and IEC 60601-2-34 standards. Therefore, there is a potential risk of electric shock. Health consequences may include electrical shock, burns, and cardiac arrhythmia.