Reported Aug 6, 2025 · Ongoing
Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35
Phasor Health, LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Phasor Health, LLC · 8944 Kirby Dr · Houston, TX · 77054-2829
Is this the product you have?
- Phasor Drill
- REF
2.70S3.20L4.505.306.35
- Lot codes and dates
- UDI-DI
B775270S0B775320L0B7754500B7755300B7756350
- Lot
24100245024100253024100945024100953024101645024101653024102345024102353024103045024103053024112027S24121827S24122727S25012227S24082832L24102332L241227450241220635250122635241227530250122530
- UDI-DI
- Amount recalled
- 1,064
Where it was sold
Why it was recalled
Drill for use in neurosurgical procedures may have a reverse battery orientation, which may increase surgical time.
This product could cause temporary or reversible health problems — stop using it.