Reported Jun 10, 2026 · Ongoing
ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system
Siemens Medical Solutions USA, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Siemens Medical Solutions USA, Inc · 40 Liberty Blvd · Malvern, PA · 19355-1418
Is this the product you have?
- ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system
- Lot codes and dates
- UDI4056869046877
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Why it was recalled
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.