Reported Aug 26, 2015 · Terminated
American Science & Engineering Inc,Gemini Cabinet X-ray system.
American Science & Engineering, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: American Science & Engineering, Inc. · 829 Middlesex Tpke · Billerica, MA · 01821-3907
Is this the product you have?
- American Science & Engineering Inc,Gemini Cabinet X-ray system.
- Lot codes and dates
- Gemini systems
- Amount recalled
- 125 systems
Where it was sold
Why it was recalled
During factory testing, we discovered that Gemini systems could, in rare circumstances, experience a fault that leads to a state in which X-rays are being produced but the X-ray indicator lights (located on the operator's console and on both ends of the tunnel) are not illuminated.
This product could cause temporary or reversible health problems — stop using it.