American Science & Engineering Inc,Gemini Cabinet X-ray system.

American Science & Engineering, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: American Science & Engineering, Inc. · 829 Middlesex Tpke · Billerica, MA · 01821-3907

Is this the product you have?

  • American Science & Engineering Inc,Gemini Cabinet X-ray system.
Lot codes and dates
  • Gemini systems
Amount recalled
125 systems

Where it was sold

Why it was recalled

During factory testing, we discovered that Gemini systems could, in rare circumstances, experience a fault that leads to a state in which X-rays are being produced but the X-ray indicator lights (located on the operator's console and on both ends of the tunnel) are not illuminated.