Reported Jul 26, 2023 · Ongoing
UFSK Treatment chair 500 XLE- Treatment chair for positioning and support of the patient during diagnostic and…
UFSK - International OSYS GmbH
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: UFSK - International OSYS Gmbh · Kirchhoffstr. 1 · Ratisbon, N/A · N/A
Is this the product you have?
- UFSK Treatment chair 500 XLE- Treatment chair for positioning and support of the patient during diagnostic and therapeutic treatments of temporary to short-term duration, especially in ophthalmology Art.no. / REF no.: 15000004 Art.no. / REF no.: 15000006
- Lot codes and dates
- Art.no. / REF no.: 15000004
- Serial number range2000- 2999 Art.no. / REF no.: 15000006
- Serial number range3000-4506
- Amount recalled
- 239 chairs US
Where it was sold
- Nationwide
Why it was recalled
Service life of the electrical seat part drive of the treatment chair may be shortened unexpectedly after several years of use due to a potential wear-related reduction in the self-locking of the drive and may remain undetected. A worsened self-locking of the seat drive can lead to overriding the drive s end position when positioning the patient with the seat part in an inclined position
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new UFSK - International OSYS GmbH recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.