Reported Aug 22, 2012 · Terminated
Sunquest Laboratory SMART & Sunquest Laboratory SMART Select
Sunquest Information Systems, Inc.
Low riskThis product is unlikely to cause harm, but it breaks a safety or labelling rule.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Sunquest Information Systems, Inc. · 250 S Williams Blvd · Tucson, AZ · 85711-4472
Is this the product you have?
- Sunquest Laboratory SMART & Sunquest Laboratory SMART Select
- Lot codes and dates
- Version 6.3.0 and Later
- Amount recalled
- 103 sites
Where it was sold
- Nationwide
Why it was recalled
In Online Entry (OEx) there are two scenarios where a Test result from one container may file to a Test on another container. In order for this to occur: 1) Site parameter SIT 1,2,1,26 (3,217) [Restrict resulting of tests to owner container (<Y>/N)]=Y. 2) The Tests are ordered on the same accession number and 3) The Test codes are defined to have the same upload code as either replicate or