Reported Aug 29, 2012 · Terminated

syngo Dynamics Version 9.5

Siemens Medical Solutions, USA, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Medical Solutions, USA, Inc · 400 Morgan Rd · Ann Arbor, MI · 48108

Is this the product you have?

  • syngo Dynamics Version 9.5
Lot codes and dates
  • 10091862 and 10091863
Amount recalled
30

Where it was sold

Why it was recalled

If multiple uncompressed images/clips are sent to the syngo Dynamics server, around the same time, it is possible for an image from a patient to be placed in another patient's study.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Medical Solutions, USA, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jul 15, 2013 · Data as of Sep 19, 5:24 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.