Reported Aug 29, 2012 · Terminated

The Instructions For Use for the following: Stryker TRIO+ 4.5 x 45 mm STANDARD POST SCREW Stryker TRIO+ 4.5 x 50 mm…

Stryker Spine

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Stryker Spine · 2 Pearl Ct · Allendale, NJ · 07401-1611

Is this the product you have?

  • The
  • Instructions For Use for the followingStryker TRIO+ 4.5 x 45 mm STANDARD POST SCREW Stryker TRIO+ 4.5 x 50 mm STANDARD POST SCREW Stryker TRIO+ 4.5 x 55 mm STANDARD POST SCREW Stryker TRIO+ 5.5 x 45 mm STANDARD POST SCREW Stryker TRIO+ 5.5 x 60
  • mm STANDARD POST SCREW Manufactured bySTRYKER Spine SAS, Z.I. Marticot - 33610 - Cestas-France +33 (0)5.57.97.06.30
  • http//www.stryker
  • com Distributed in the USA byStryker Spine, 2 Pearl Ct., Allendale, NJ 07401-1677 USA +1-201-750-8000. The system is intended for posterior, non-cervical pedicle and non-pedicle pedicle fixation of the spine
  • It is indicated fordegenerative disc disease (DDD), spondylolisthesis, fracture, dislocation, spinal stenosis, scoliosis kyphosis, lordosis, spinal tumor, pseudarthrosis, and failed previous fusion
Lot codes and dates
  • Code information
  • listed in the following formatManufacturer Part Number/Prodcut Description/Corresponding IFU Part Number/Affected Lot #s
  • are as follows48900445/Trio Standard Post Screw 5.5x55mm/#NOLI135B03Rev02/57281 and 57201 48900450/Trio Standard Post Screw 5.5x60mm/#NOLI135B03Rev02/57197 48900455/Trio Standard Post Screw 4.5x45mm/#NOLI135B03Rev02/57198 48900555/Trio Standard Post Screw 4.5x50mm/#NOLI135B03Rev02/56599 48900560/Trio Standard Post Screw 4.5x55mm/#NOLI135B03Rev02/57192
Amount recalled
40 units.

Where it was sold

Why it was recalled

The front cover of Instructions For Use (IFU) #NOLI135B03Rev02 was improperly marked with a "Sterile" logo in the lower right corner. The products these IFUs were packed with are sold as "Non-Sterile" and require sterilization before use in surgery.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Stryker Spine recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Feb 27, 2013 · Data as of Sep 19, 5:24 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.