Reported Aug 3, 2016 · Terminated
GE Inspection Technologies, LP - Phoenix v|tome|x m240 cabinet x-ray systems. A microfocus CT system used primarily for…
GE Inspection Technologies, LP
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: GE Inspection Technologies, LP · 50 Industrial Park Rd · Lewistown, PA · 17044-9312
Is this the product you have?
- GE Inspection Technologies, LP - Phoenix v|tome|x m240 cabinet x-ray systems. A microfocus CT system used primarily for 3D metrology and analysis.
- Lot codes and dates
- SN - PA2384
- Amount recalled
- 4
Where it was sold
Why it was recalled
Radiation emissions which exceeded the limit in the performance standard were detected during maintenance of one system. The identified cause was insufficient attachment of shielding in the door. Jostling during delivery was identified as the probable cause of this failure to comply with the performance standard.
This product could cause temporary or reversible health problems — stop using it.