Reported Aug 2, 2023 · Ongoing
Ringed DxI Reaction Vessels (RVs)
Beckman Coulter, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Beckman Coulter, Inc. · 1000 Lake Hazeltine Dr · Chaska, MN · 55318-1037
Is this the product you have?
- Ringed DxI Reaction Vessels (RVs)
- Lot codes and dates
- UDI15099590229979
- Catalog Number386167
- Lot Numbers
L17855111L17798598L18046334
- Amount recalled
- 213 cases of 10,000 RVs (2,130,000 RVs)
Where it was sold
Why it was recalled
Beckman Coulter, Inc. is recalling Ringed DxI Reaction Vessels (RVs) as part of Unicel DxI Immunoassay System (600, 800 series). The reason for the recall is that the firm has identified 3 lots of RVs that failed an internally used service test to evaluate washing efficiency. Increased relative light units (RLU) suspected to be caused by the resulting rough interior surface may lead to artificially falsely elevated results for susceptible tests, in particular, sandwich assays including, but not limited to, hsTnI, Hybritech p2PSA, IPTH/IPTH IO.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Beckman Coulter, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.