Reported Jul 10, 2024 · Ongoing
WOM Tube Set for lrrlgatlon, single-use- Intended to transport irrigation fluid to the patient in endoscopic surgeries.…
WOM World Of Medicine AG
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: WOM World Of Medicine AG · Alte Poststr. 11 · Ludwigsstadt, N/A · N/A
Is this the product you have?
- WOM Tube Set for lrrlgatlon, single-use- Intended to transport irrigation fluid to the patient in endoscopic surgeries. Distr. by Endoscopy Smith & Nephew Inc
- REF NumbersT0449-01
- Lot codes and dates
- UDI04056702003678
- Lot Numbers
40309024030903
- Amount recalled
- 1700 units
Where it was sold
- Nationwide
Why it was recalled
Incorrect transponder data was written on the RFID tags triggers the error message E03, Tube Set lnvalid, may result in a delay of surgery