Reported Jun 27, 2018 · Terminated

Perseus A500 Anesthesia Machine; Cat. no. MK06000 Product V Product Usage: Intended for use in anesthetizing adults,…

Draeger Medical, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Draeger Medical, Inc. · 3135 Quarry Rd · Telford, PA · 18969-1042

Is this the product you have?

  • Perseus A500 Anesthesia Machine; Cat. no
  • MK06000 Product V Product UsageIntended for use in anesthetizing adults, children, and neonates and can be used for automatic and manual ventilation, pressuresupported spontaneous breathing, and spontaneous breathing. Perseus is equipped with airway monitoring, gas measurement and device monitoring, O2 insufflation, and an anesthetic gas receiving system. Anesthesia is achieved through a mixture of pure oxygen and Air (medical compressed air) or pure oxygen and nitrous oxide, with the addition of volatile anesthetic agents. Ventilation is accomplished on the patient through a laryngeal mask airway, a full-face mask, or an endotracheal tube. The integrated breathing system can be used with partial rebreathing (low-flow or minimum-flow). A non-rebreathing system such as the Kuhn or Medec Water System may be used at the external fresh-gas outlet (optional). Perseus is specified for inhalational anesthesia and/or patient ventilation in accordance with the intended use during surgical or diagnostic interventions
Lot codes and dates
    • ASKM-0197
    • ASKN-0006
    • ASKN-0007
    • ASKN-0008
    • ASKN-0009
    • ASKN-0013
    • ASKN-0014
    • ASKN-0015
    • ASKN-0021
    • ASKN-0026
    • ASKN-0027
    • ASKN-0028
    • ASKN-0029
    • ASKN-0030
    • ASKN-0031
    • ASKN-0034
    • ASKN-0035
    • ASKN-0036
    • ASKN-0037
    • ASKN-0038
    • ASKN-0039
    • ASKN-0040
    • ASKN-0042
    • ASKN-0045
    • ASKN-0046
    • ASKN-0048
    • ASKN-0049
    • ASKN-0050
    • ASKN-0053
    • ASKN-0054
    • ASKN-0055
    • ASLA-0009
    • ASLA-0010
Amount recalled
33

This recall covers 5 products

The agency filed one notice per product. You are reading one of them; the other 4 are below.

Where it was sold

Why it was recalled

The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes include death of the patient.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Draeger Medical, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated May 12, 2020 · Data as of Sep 19, 6:15 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.