Reported Aug 10, 2016 · Ongoing
Olympus Scientific Solutions Americas Corporation (OSSA) Model: Analytical X-Ray System X -Ray Fluorescence. Intended…
Olympus Scientific Solutions Americas
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Olympus Scientific Solutions Americas · 48 Woerd Ave · Waltham, MA · 02453-3824
Is this the product you have?
- Olympus Scientific Solutions Americas Corporation (OSSA)
- ModelAnalytical X-Ray System X -Ray Fluorescence. Intended to emit
- ionizing radiation Delta with models
DC-2000DC-4000DC-6000DI-2000DP-2000DP-4000DP-4050DP- 6000DPO-4000DPO-6000DS-2000DS-4000DS-6000DS-6500DeltaDelta 50Delta Classic PlusDelta ElementDelta PremiumDelta Professionaland Delta-Pl us
- Lot codes and dates
- All Models of the Delta XRF Analyzer with Software version 2.5.17.1 and higher
- Amount recalled
- 2, 210 units distributed in the USA
Where it was sold
- Nationwide
Why it was recalled
Trigger lock option does not always prevent x-ray generation. Specifically if the unit is in test mode or repeat test mode then the trigger lock may not prevent x-ray initiation when the trigger is pulled. The correction is via repair by a software update.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Olympus Scientific Solutions Americas recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.