Reported Aug 3, 2016 · Terminated

Container fill and transfer sets, 250 mL, 500 mL, 1000 mL, 2000 mL, 3000 mL, or 4000 mL.

The Metrix Company

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: The Metrix Company · 4400 Chavenelle Rd · Dubuque, IA · 52002-2655

Is this the product you have?

    • Container fill and transfer sets
    • 250 mL
    • 500 mL
    • 1000 mL
    • 2000 mL
    • 3000 mL
    • or 4000 mL
Lot codes and dates
  • All lots are being recalled. Lot code consists of the five digit product code followed by a dash followed by a unique five character numeric or alphanumeric sequence
  • Affected product codes for this product
    • 58781
    • 58785
    • 62005
    • 62010
    • 62015
    • 62025
    • 62030
    • 62035
    • 64005
    • 64010
    • 64015
    • 64025
    • 64030
    • 64035
    • 64781
    • 65005
    • 65010
    • 65015
    • 65025
    • 65030
    • 65035
    • 66005
    • 66010
    • 66015
    • 66020
    • 66025
    • 66030
    • 66035
    • 66305
    • 66310
    • 66315
    • 66325
    • 66330
    • 66335
    • 66705
    • 66710
    • 66715
    • 66725
    • 66730
    • 66735
    • 66910
    • 66915
    • 66925
    • 66930
    • 66935
    • 67005
    • 67010
    • 67015
    • 67025
    • 67030
    • 67035
    • 69005
    • 69010
    • 69015
    • 69025
    • 69030
    • 69035
Amount recalled
137,598 cases

This recall covers 8 products

The agency filed one notice per product. You are reading one of them; the other 7 are below.

Where it was sold

Why it was recalled

An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new The Metrix Company recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Apr 11, 2023 · Data as of Sep 19, 5:19 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.