Reported Jul 10, 2024 · Ongoing
Abbott Liberta RC, Implantable Pulse Generator, REF: 62400, STERILE, RxONLY
Abbott Medical
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Abbott Medical · 6901 Preston Rd · Plano, TX · 75024-2508
Is this the product you have?
- Abbott Liberta RC, Implantable Pulse Generator
- REF62400, STERILE, RxONLY
- Lot codes and dates
- UDI-DI05415067047663/
- Serial Numbers
302 of 302 codes
201964932019649520196496201964972019649820196500201965022019650320196504201965052019650620196507201965082019650920196511201965122019651320196514201965152019651620196517202073972020739820207399202074002020740120207402202074032020740420207406202074072020740820207409202074102020741120207412202074132020741620207417202074192020742020207421202074222020742320207424202074252020742720207428202074292020743020207431202074322020743320207434202074352020743620207437202074382020744120207442
- Amount recalled
- 302 units
Where it was sold
Why it was recalled
Deep brain stimulation system will first turn off after approximately 50 days after the system is activated and subsequently, this same action will recur approximately every 50 days thereafter. This unexpected lost of stimulation may result in loss of therapy.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Abbott Medical recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.