Reported Aug 5, 2015 · Terminated
Prism Medical P-600 Portable Ceiling Lift; PN 303090. Patient lift with weight specification of 600 lbs (273 kg).…
ErgoSafe Products, LLC (DBA) Prism Medical
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: ErgoSafe Products, LLC (DBA) Prism Medical · 10888 Metro Ct · Maryland Heights, MO · 63043-2413
Is this the product you have?
- Prism Medical P-600 Portable Ceiling Lift; PN 303090. Patient lift with weight specification of 600 lbs (273 kg)
- Product UsageThe P-600 lift is a lifting aid used by health care professionals and those providing care in the home to lift, position, and transfer clients or a disabled family member. The ceiling lift makes it possible to move mobility impaired individuals with minimal strain or risk to the caregiver, while providing complete safety, dignity, and comfort for the client or family member
- Lot codes and dates
- All serial numbers.
- Amount recalled
- 409 lifts
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
- Nationwide
Why it was recalled
There is a potential the sling loops may not stay attached to the carry bar hooks.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new packaged foods recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.