Reported Sep 5, 2012 · Terminated
PHADIA Varelisa Cardiolipin IgM Cardiolipin autoantibody immunological test system (IgM class antibodies)
Phadia US Inc.
Low riskThis product is unlikely to cause harm, but it breaks a safety or labelling rule.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Phadia US Inc · 4169 Commercial Ave · Portage, MI · 49002-9701
Is this the product you have?
- PHADIA Varelisa Cardiolipin IgM Cardiolipin autoantibody immunological test system (IgM class antibodies)
- Lot codes and dates
- Article Number 15696 Lot number #0067
- Amount recalled
- 837 kits (96 assays per kit), total of 80,352 individual assays
Where it was sold
- Nationwide
Why it was recalled
Customers were not following the labeled assay procedure. This was due to a letter issued to them in 2004 from Phadia GmbH, which indicated that customers could omit the pre-wash step. The omission of the pre-wash step, when combined with a change in material source in kit lot #0067, caused elevated results.
This product is unlikely to cause harm, but it breaks a safety or labelling rule.