Reported Jun 10, 2026 · Ongoing

Philips Azurion 3M15, Model Numbers: 722064, 722222 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6…

Philips Medical Systems Nederland B.v.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Veenpluis 6 · Best · N/A

Is this the product you have?

  • Philips Azurion 3M15
  • Model Numbers722064, 722222 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
Lot codes and dates
    • UDI (01)00884838085282(21)209
    • (01)00884838099210(21)100
    • (01)00884838099210(21)254
    • (01)00884838099210(21)256
    • (01)00884838099210(21)257
    • (01)00884838099210(21)265
    • (01)00884838099210(21)255
    • (01)00884838099210(21)99
    • (01)00884838085282(21)260
    • (01)00884838085282(21)261
    • (01)00884838099210(21)69
    • (01)00884838099210(21)12
    • (01)00884838085282(21)314
    • (01)00884838085282(21)259
    • (01)00884838099210(21)15
    • (01)00884838099210(21)19
    • (01)00884838099210(21)23
    • (01)00884838099210(21)17
    • (01)00884838085282(21)258
    • (01)00884838099210(21)55
    • (01)00884838099210(21)45
    • (01)00884838085282(21)270
    • (01)00884838099210(21)187
    • (01)00884838099210(21)223
    • (01)00884838099210(21)44
    • (01)00884838099210(21)113
    • (01)00884838099210(21)112
Amount recalled
27 units

This recall covers 8 products

The agency filed one notice per product. You are reading one of them; the other 7 are below.

Where it was sold

  • Nationwide

Why it was recalled

During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Philips Medical Systems Nederland B.v. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:37 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.