Reported Aug 27, 2014 · Terminated

PERSONA (TM) The Personalized Knee System Tibial General Instrument Tray, non-sterile The intended use of the…

Zimmer, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Zimmer, Inc. · 1800 W Center St · Warsaw, IN · 46580-2304

Is this the product you have?

  • PERSONA (TM) The Personalized Knee System Tibial General Instrument Tray, non-sterile The intended use of the sterilization case is to organize, store and protect its contents from damage consistent with typical hospital or surgical center circulation (i.e. between central services and the operating room) and during transport between a Zimmer distributorship and a health care facility by Zimmer distributorship personnel.
Lot codes and dates
  • Item Number 00-5908-015-00
  • Lot Numbers56519125 & 56518861
Amount recalled
190 units

This recall covers 16 products

The agency filed one notice per product. You are reading one of them; the other 15 are below.

3 more are in the archive.

Where it was sold

Why it was recalled

The firm discovered that the current Personal Tibial/General Fixed Bracket Trays may not have the specific geometry to accommodate the tibial drill and stop guide with proper position/orientation. The drill bracket and drill guide are not the correct geometry to allow for a proper fit in the Persona Fixed Case Kit which could lead to improper sterilization of the tibial drill and stop guide.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Zimmer, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jun 9, 2015 · Data as of Sep 19, 6:03 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.