Reported Aug 3, 2016 · Terminated

Babytherm 8004/8010, Infant Warmer, Class II device, 510 K K971198, Catalog # FR00105 (Babytherm 8010), 2M30404…

Draeger Medical, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Draeger Medical, Inc. · 3135 Quarry Rd · Telford, PA · 18969-1042

Is this the product you have?

  • Babytherm 8004/8010, Infant Warmer, Class II device, 510 K K971198, Catalog # FR00105 (Babytherm 8010), 2M30404 (Babytherm 8004), non sterile, software controlled
  • Product UsageBabytherm 8004/8010 is an open care infant warmer system for warming premature babies, neonates and infants with a body weight of up to 8 kg (17.6 lbs). The device is intended for use in delivery rooms, operating rooms, maternity and obstetric units, neonatal and pediatric wards, neonatal and pediatric intensive care units
  • Therapy and nursing uses- As a resuscitation table and in birthing rooms, obstetric units, and NICUs - As infant warmer bassinet during normal and intensive care - For infant re-warming and compensation of heat loss - For thermal stabilizing of infants with mattress heater (Babytherm 8010 only) - For gentle warming therapy using the combination of radiant and mattress heater - Cooling infants with fever (temperature of mattress below infant body temperature, Babytherm 8010 only) - With adjustable bed tilt for treatment of infants in Trendelenburg and reverse Trendelenburg position - For weaning infants from incubators - For O2 and nebulizer therapy (with appropriate accessories) - For lowering an infant s bilirubin levels when using the phototherapy option
Lot codes and dates
  • Serial NumbersARLN-xxxx
  • to ASFE-xxxx Catalog Numbers
    • FR00105 (Babytherm 8010)
    • 2M30404 (Babytherm 8004)
Amount recalled
1,599 units

Where it was sold

Why it was recalled

Risk associated with operating the device in skin temperature mode with temperature sensors attached but no patient connected.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Draeger Medical, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Dec 7, 2016 · Data as of Sep 19, 6:00 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.