Reported Oct 9, 2013 · Terminated
Lactate Membrane units for E7077 Electrode Radiometer Medical ApS. The intended use is in vitro testing of samples of…
Radiometer America Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Radiometer America Inc · 810 Sharon Dr · Westlake, OH · 44145-1521
Is this the product you have?
- Lactate Membrane units for E7077 Electrode Radiometer Medical ApS. The intended use is in vitro testing of samples of whole blood.
- Lot codes and dates
- Part Number942-066 R1741 to R1869
- Amount recalled
- 13,632 boxes/4 membranes
Where it was sold
Why it was recalled
RADIOMETER became aware that some membranes may have enzyme residue on the outer membrane. The enzyme residue may cause an initial negative bias on the reported Lactate result upon replacement of the Lactate membrane.
This product could cause temporary or reversible health problems — stop using it.