Reported Aug 25, 2021 · Terminated

Oxygen Sensor Model K54019-2 used in the: G185 & G210 Incubators and G603 Work Chambers Oxygen Sensors supplied for use…

CooperSurgical, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: CooperSurgical, Inc. · 95 Corporate Dr · Trumbull, CT · 06611-1350

Is this the product you have?

  • Oxygen
  • Sensor Model K54019-2 used in theG185 & G210 Incubators and G603 Work Chambers Oxygen Sensors supplied for use in the following incubator and work chamber products; G185 Standard 110V, G185 SensorTech 110V, G185 Standard 230V, G185 SensorTech 230V, G210 InviCell Standard, G210 InviCell Standard US, G210 Invicell Plus Optional pH and External Monitoring, G603 Work Chamber 115V, & G603 Work Chamber 230V. [CooperSurgical part numbers K22400-110, K22200-110, K22400-230, K22200-230, K59500, K59700, K60000, 3.641.831.153, & 3.641.832.153]
Lot codes and dates
  • Serial Numbers

    430 of 430 codes

    • 531540
    • 531541
    • 531542
    • 531543
    • 531544
    • 531545
    • 531546
    • 531547
    • 531548
    • 531549
    • 531550
    • 531551
    • 531552
    • 531553
    • 531554
    • 531555
    • 531556
    • 531557
    • 531558
    • 531559
    • 531560
    • 531561
    • 531562
    • 531563
    • 531564
    • 531565
    • 531566
    • 531567
    • 531568
    • 531569
    • 531570
    • 531571
    • 531572
    • 531573
    • 531574
    • 531575
    • 531576
    • 531577
    • 531578
    • 531579
    • 531580
    • 531581
    • 531582
    • 531583
    • 531584
    • 531585
    • 531586
    • 531587
    • 531588
    • 531589
    • 531590
    • 531591
    • 531597
    • 531598
    • 531599
    • 531600
    • 531601
    • 531602
    • 531603
    • 531604
Amount recalled
421 units

Where it was sold

Why it was recalled

Units are displaying output readings that differ from the measurements taken using a Gas Analyzer. When the display is reading a target setting of 5%, an external meter is reading between 3% and 12% and the Incubator does not alarm. May result in loss of embryo during storage.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new CooperSurgical, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Nov 21, 2022 · Data as of Sep 19, 6:21 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.