Reported Aug 25, 2021 · Ongoing
Cell Marque Antibodies, PSA (ER-PR8) Mouse Monoclonal Antibody, REF: 324M-15, 324M-16
Cell Marque Corporation
Low riskThis product is unlikely to cause harm, but it breaks a safety or labelling rule.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Cell Marque Corporation · 6600 Sierra College Blvd · Rocklin, CA · 95677-4306
Is this the product you have?
- Cell Marque Antibodies, PSA (ER-PR8) Mouse Monoclonal Antibody
- REF
324M-15324M-16
- Lot codes and dates
- Item/Lot
324M-15/0000116695324M-16/0000116696
- Item/Lot
- Amount recalled
- 20
Where it was sold
Why it was recalled
Mouse Monoclonal Antibody labelled with an incorrect expiry date of 31 Dec 2023. The correct expiry date is 31 Dec 2022. If used beyond expiry, the material performance is likely to decrease and may be observed as weak or no staining, and there is a potential risk that a false negative result could be recorded and may lead to delay of diagnosis due to retest.
This product is unlikely to cause harm, but it breaks a safety or labelling rule.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Cell Marque Corporation recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.