Reported Jul 17, 2024 · Completed

MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display

Barco N.V.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Barco N.V. · President Kennedypark 35 · Kortrijk, N/A · N/A

Is this the product you have?

    • MNA-420 ENC HDMI
    • Material Number K9303311
    • cathode ray tube display
Lot codes and dates
  • serial numbers

    88 of 88 codes

    • 2530986299
    • 2530986300
    • 2530986301
    • 2530986302
    • 2530986303
    • 2530986304
    • 2530986305
    • 2530986306
    • 2530986307
    • 2530986308
    • 2530986309
    • 2530986310
    • 2530986311
    • 2530986312
    • 2530986313
    • 2530986314
    • 2530986315
    • 2530986316
    • 2530986317
    • 2530986318
    • 2530986319
    • 2530986320
    • 2530986321
    • 2530986322
    • 2530986323
    • 2530986324
    • 2530986325
    • 2530986326
    • 2530986327
    • 2531041594
    • 2531041591
    • 2531041578
    • 2531041585
    • 2531041595
    • 2531041596
    • 2531041586
    • 2531041588
    • 2531041587
    • 2531041577
    • 2531041584
    • 2531041581
    • 2531041612
    • 2531041608
    • 2531041600
    • 2531041601
    • 2531041604
    • 2531041599
    • 2531041602
    • 2531041605
    • 2531041606
    • 2531041607
    • 2531041609
    • 2531041610
    • 2531041580
    • 2531041583
    • 2531041598
    • 2531041593
    • 2531041590
    • 2531041614
    • 2531041582
Amount recalled
88 units

Where it was sold

Why it was recalled

We have identified a potential issue with the SFP+ module supplied by Molex (Part Number: B620050) when connected to the ETH0 slot of either the MNA-6x0 ENC HDMI or MNA-4x0 ENC HDMI. In some instances, this connection may render the first HDMI channel inaccessible, making both the channel and the respective MNA ENC unit unusable. Additionally, readings from the module become unavailable when this issue arises. Similarly, when this Molex module is connected to the ETH0 slot of an MNA-620 ENC ANA, readings may also become inaccessible.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Barco N.V. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 4:54 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.