Reported Sep 19, 2012 · Terminated

Aesculap S4 Element 4.5mm Polyaxial Screws (ST240T, ST241T, ST242T, ST243T, ST244T, and ST245T). The S4 Spinal System…

Aesculap Implant Systems LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Aesculap Implant Systems LLC · 3773 Corporate Pkwy · Center Valley, PA · 18034-8217

Is this the product you have?

  • Aesculap S4 Element 4.5mm Polyaxial Screws (ST240T, ST241T, ST242T, ST243T, ST244T, and ST245T). The S4 Spinal System is intended for anterior/anterolateral and posterior, non-cervical pedicle and non-pedicle fixation. Fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion using autograft or allograft. The device is indicated for treatment of the following acute and
  • chronic instabilities or deformities
    • 1 - degenerative disc disease
    • 2 - spondylolisthesis
    • 3 - trauma
    • 4 - spinal stenosis
    • 5 - deformities or curvatures
    • 6 - tumor
    • 7 - pseudoarthrosis
    • and 8 - failed previous fusion
Lot codes and dates
  • All distributed lots and serial numbers
Amount recalled
5

Where it was sold

Why it was recalled

There is a possibility that the screw channel that holds the rod in place may be out of tolerance. When the depth of the channel is not deep enough, the rod may not be able to provide ample compression on the insert preventing the screw head from locking down to the bone screw.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Aesculap Implant Systems LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Mar 18, 2014 · Data as of Sep 19, 5:24 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.